Configure the separations module
July 29, 2013, 715003794 Rev. B 91
6. Press the stuvwx screen key.
7. Press the v screen key.
8. Press the abcdef screen key.
9. Press the e screen key.
10. Press Enter to store the entry.
Configure the separations module
Before you can operate the separations module, you need to configure the
separations module for standalone or remote control and also for various
operating conditions. Configure the separations module using the
Configuration screen whenever you want to perform one of these tasks:
• Set up the separations module for the first time
• Change the configuration of the separations module
• Add, change, or delete a user
• Reset the clock
Press the Configure screen key in the Main screen to display the
Configuration screen.
Set configuration parameters
Select the following parameters, as necessary, for your configuration:
• Syringe size, Loop size – Specify the size of the syringe and sample loop.
These values are preset at the factory. Change them only when you
install a different syringe and/or sample loop.
• Verify vial presence – When checked (the default setting), the software
notifies you when a vial is not present at a specified carousel position.
• Verify carousel placement – When checked (the default setting), the
software notifies you when a carousel is mounted in the wrong position
at the time when an injection is attempted.
• System – Selects the operating mode and network address, when
required. See “Configure the operating mode” on page 98.
• Printer – Selects the type of printer and the baud (data transfer) rate.