Maintain the sample management system
July 29, 2013, 715003794 Rev. B 197
5. Insert the replacement in-line filter element into the in-line filter
housing (see the figure “Replacing the in-line filter:” on page 197).
Replacing the in-line filter:
6. Reconnect the in-line filter inlet and outlet housings.
7. Retighten the compression screw into the inlet housing.
8. Flush the solvent management system with mobile phase at 1 mL/min
for 10 minutes (see “Power-off the separations module” on page 114).
9. Inspect the in-line filter for leaks, and tighten the fittings as required.
Maintain the sample management system
Perform the procedures in this section when you determine there is a problem
with a sample management system component.
Inlet housing
In-line filter
element
Outlet housing