2 Install the e2695 Separations Module
78 July 29, 2013, 715003794 Rev. B
As with IEEE-488 control, triggering of Waters Ethernet-controlled devices,
such as 2489, 2998, 2414, and 2424 detectors, occurs over the Ethernet cable,
and so requires no external I/O cables.
Exception: Waters 2475 Fluorescence Detector, v2.01 requires an external I/O
trigger cable under Ethernet control.
Configure the Ethernet port
On the Configuration screen, select Auto/Ethernet for the separations module.
See “Configure the operating mode” on page 98 for information about
configuring the Ethernet port.
Add hardware options
Depending on your applications, you can install the following options on the
separations module:
• Larger sample loop
• Column heater
• Column heater/cooler
• Larger or smaller syringe
• Column selection valve
Install an optional sample loop
To inject sample volumes greater than 100 µL, install an optional sample loop.
To do so, you must perform these tasks:
• Remove the right-hand side panel
• Install the sample loop
• Refit the right-hand side panel
Required materials
• Screwdriver, #2 Phillips
• Wrenches, 5/16-inch, two
• Wrench, adjustable
• Sample Loop Option Kit