Monitor your HPLC system
July 29, 2013, 715003794 Rev. B 119
Load the carousel
To load the carousel:
1. Open the carousel door if it is closed.
2. Press the Next screen key (or select the desired carousel screen key) to
position the carousel turntable for loading the appropriate carousel.
3. Load the carousel into the sample compartment, and close the carousel
door.
Tip: If you inadvertently load a carousel in the wrong location when the
“Verify Carousel Placement” parameter in the Configuration screen is
selected, an error message appears when you make an injection from
that carousel.
Add new sample vials during a run
If you need to add new sample vials while the instrument is processing a
sample set, use these guidelines to minimize interruptions during the run:
Have the new sample vial(s) ready so that you minimize the amount of
time that the sample compartment door is open.
Wait until the run time displayed on the front panel is greater than 30
seconds before adding the sample(s).
Do not add any sample(s) during the last 30 seconds of the run.
Ensure that the sample compartment door is closed at the start of each
injection when injection commands (selecting vial, washing the needle,
drawing sample, and so forth) are executing.
Monitor your HPLC system
You can use the Status screen to monitor the current status of your HPLC
system during a run. Press Menu/Status to display the first page of the Status
screen. The content and layout of the Status screen vary with the options
installed in the separations module and with the mode of operation.