EasyManua.ls Logo

Waters Alliance e2695 - Set Autosampler Parameter Values

Waters Alliance e2695
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Set separation method parameters
July 29, 2013, 715003794 Rev. B 159
Set autosampler parameter values
To enter parameters in the Autosampler screen:
1. Press the Next or Prev screen key to display the Autosampler screen.
Autosampler screen:
Syringe Draw
Rate
Custom
Selects one of three pre-set syringe
draw rates to accommodate viscous
samples. The rate changes with the
size of the installed syringe.
Specifies a user-entered custom draw
rate value.
Fast (5.0 µL/sec)
1
Normal (2.5 µL/sec)
Slow (1.0 µL/sec)
Custom (1.00 to
5.00 in 0.01 µL/sec
increments)
Sample Draw
Depth
2
Adjusts the depth of the needle tip to
accommodate for sedimented samples
or nonstandard vials. A value of 0
corresponds to the bottom of the vial.
2
0 to 20 in 1 mm
increments
2
1. These draw rates are for a 250-µL syringe. The values automatically change to reflect the
syringe size entered in the Configuration screen.
2. See “Vials and low volume inserts” on page 286, for a list of sample draw depths required for
use with separations module vials and Low Volume Inserts.
Sample parameters: (Continued)
Parameter Function Value range

Table of Contents

Related product manuals