EasyManua.ls Logo

Waters Alliance e2695 - Change from a Buffered Solvent to an Organic Solvent; Power-Off the Separations Module

Waters Alliance e2695
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Prepare the e2695 Separations Module for Operation
114 July 29, 2013, 715003794 Rev. B
Change from a buffered solvent to an organic solvent
To change from a buffered solvent to a totally organic solvent in the
separations module:
1. Remove the solvent line from the reservoir containing the buffered
solvent.
2. Specifying a flow rate of 2.000 mL/min, perform a wet prime until air
begins to enter the line (see “Wet prime the system” on page 129).
3. Place the previously removed solvent line into filtered HPLC-grade
water, and perform a wet prime for 0.5 min at 5 mL/min (see “Wet prime
the system” on page 129).
4.
Equilibrate the solvent in the degasser chamber by setting a flow rate of
0.000 mL/min for 5 min
(see “Equilibrate solvents in the in-line vacuum
degasser” on page 130).
5. Purge the injector for a minimum of 6 loop volumes (see “Purge the
injector” on page 131).
6. Remove the solvent line from the HPLC-grade water.
7. Fill a clean solvent reservoir with the organic solvent and place the
solvent line in the reservoir.
8. Perform the procedure, above, for preparing an idle or powered-off
separations module for operation.
Power-off the separations module
Before you power-off the separations module, remove any buffered mobile
phase present in the flow path (see “Flushing the flow path” below).
Caution: To avoid salts from precipitating in the separations module,
use an intermediate solvent such as water when you change from
buffers to high-organic-content solvents (see “Solvent miscibility” on
page 297).

Table of Contents

Related product manuals