Prepare the sample management system for operation
July 29, 2013, 715003794 Rev. B 107
6. Push on the syringe plunger to force seal-wash solvent through the
system.
7. When seal-wash solvent flows out the seal-wash waste tube, press the
Halt screen key.
8. Reinstall the inlet filter, and place the seal-wash inlet tube in the
seal-wash reservoir.
9. Press Close to return to the Diagnostics screen.
Prime the solvent management system
Prime the solvent management system using one of these priming methods:
Dry prime, when the solvent lines are dry
Wet prime, when you want to change between miscible solvents
Follow the procedure in “Dry prime the system” on page 126 or “Wet prime the
system” on page 129, as appropriate.
Once you prepared the solvent management system for operation, you must
prepare the sample management system for operation.
Prepare the sample management system for operation
Prepare the sample management system for operation after you prepared the
solvent management system (see “Prepare the solvent management system”
on page 103). The following figure shows the steps involved in preparing the
sample management system for operation.