Set detector parameters
In a standalone mode, System Controller or No Interaction, the separations
module can control one Waters 2414, 2410, or 410 Differential Refractometer
Parameter Selects the value for the action. See the table
“Action
parameters”,
below.
Action parameters:
Event type Action Value range (Param column)
Switches 1 to 4 On
Off
Toggle (Changes the
state of the switch
(open to closed, or
closed to open).
Pulse (A single pulse
with a width defined
in the param column
of the I/O Events
Table.)
No Change
0.01 to 10.00 in 0.01-min
increments (Pulse only)
Set Temperature
if sample
heater/cooler,
column heater, or
column
heater/cooler are
installed
Sample
Column heater
Column heater/cooler
4 to 40 °C in 1 °C increments
20 to 60 °C (5 °C above
ambient to 60 °C) in 1 °C
increments
4 to 65 °C (the greater of 15 °C
below ambient or 4 to 65 °C) in
1 °C increments
Alert No action N/A