EasyManua.ls Logo

Waters Alliance e2695 - Page 167

Waters Alliance e2695
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Set separation method parameters
July 29, 2013, 715003794 Rev. B 167
and as many as two UV/Vis detector channels (Waters 2487 or Waters 486
detectors). Use the IEEE-488 interface bus to connect the separations module
to the detector(s). To make I/O connections to these and other detectors, see
“I/O signal connections” on page 71.
Tip: In Ethernet mode, you set up the detector parameters via the
chromatography data software (see the detector operator’s guide for specific
information).
To enter parameters in the Detectors screen:
1. Press the Next or Prev screen key to display the Detectors screen.
Detectors screen:
2. In the Detectors screen, use the up/down arrow keys to select the
appropriate detector (for example, Absorbance Detector 2), and then
press Enter to display the detector selections.
3. Use the up/down arrow to highlight your selection, and then press Enter
to accept your selection.
The following table describes the selections in the Detectors screen.

Table of Contents

Related product manuals