Set separation method parameters
July 29, 2013, 715003794 Rev. B 165
I/O Events table:
2. Enter values in the I/O events table as appropriate, and then press Exit.
The following table describes the I/O Events table’s parameters.
I/O Events table parameters:
Parameter Function Value range
Time Determines the time after the start of
a run during which the change is to
occur. Press the Clear key to select
INIT. Conditions in the INIT line
apply when the system is initialized
to a method, while events at time
0.00 occur immediately upon an
injection.
INIT, 0.00 to
999.99 in 0.01-min
increments
Event type Sets the type of event to occur. Switches 1 through
4
Set Temp
Alert
Action Selects the action to perform with the
specified event.
See the table
“Action
parameters”,
below.