6 Create Methods, Sample Sets, and Sample Templates
164 July 29, 2013, 715003794 Rev. B
Edit the I/O Events table
In the I/O Events table, you set the timing for the following events to occur
during a run:
Changing the state of event switches
Setting the column temperature
Setting the sample compartment temperature
•Alerts
You can program up to 25 lines in the I/O Event table.
To enter event in the I/O Events table:
1. Press the Table screen key in the I/O screen to open the I/O Events table
screen.
Chart Output Defines the signal sent out on the
chart out terminals on the rear panel
of the separations module.
Flow Rate
System Pressure
Sample Loop
Pressure
%A, %B, %C, %D
Column Temp
Sample Temp
Degasser Vacuum
Primary Head
Pressure
Timed Table
(screen key)
Displays the I/O Events table. See the “Editing
the I/O Table”
discussion next.
I/O parameters: (Continued)
Parameter Function Value range