3 Prepare the e2695 Separations Module for Operation
92 July 29, 2013, 715003794 Rev. B
Tip: The baud rate is automatically set to 1200 when you set the Waters
746 Integrator as the printer type.
Date format – Selects between MDY (for example, Aug 28 07) or DMY
(28 Aug 07).
Pressure units – Selects between psi, kilopascal (kPa), and bar units.
Alarm sound – Selects the duration of alarms (continuous or a single
beep).
Press the screen keys in the Configuration screen to access other parameters
whose values you can specify.
Screen keys in the Configuration screen:
Screen key Description
User Names Displays a list of stored user names and adds, changes,
and removes them.
Detectors Scans the IEEE-488 bus and displays a list of Waters
detectors connected to it, their addresses, and their
firmware revision numbers. Active only when the
separations module is configured as the system
controller and the detectors are powered-on.
Options Displays a list of the hardware options installed on the
separations module.
Set Clock Displays the Set Clock dialog box, in which you
establish the current time and date.
Events In Specifies external signal conditions for Stop Flow and
Inject Hold (see “Configure the operating mode” on
page 98).
More Toggles between the two sets of screen keys for the
Configuration screen.
Reports Specifies the type of information that is included in
reports, and whether the reports go to a printer or
integrator.
Comp Check Specifies the criteria for the compression check.
Auto Shutdown Specifies parameters for automatic shutdown of the
separations module (see “Auto shutdown” on page 93).