4 Front Panel Control
134 July 29, 2013, 715003794 Rev. B
2. Insert the carousel in the sample compartment, as described in “Load
sample vials” on page 118.
3. Close the carousel door.
Tip: If the door is ajar, the separations module does not perform the
injection.
4. In the Status screen, enter the desired initial conditions, or select the
Method field, and then select the separation method that specifies the
conditions under which you want to perform the injection.
5. Press the Direct Function screen key.
6. Select the Inject Samples option, and then press Enter.
7. In the Inject Samples dialog box, specify the vial range, which includes
the vial positions from which you want to make injections.
Tip: Use the “.” key to separate the entries. For example, to inject
samples from vials 1 through 20, enter 1.20, then press Enter.
8. Specify the number of injections from each vial, and then press Enter.
9. Specify the run time, and then press Enter.
10. Specify the volume of each injection, and then press the OK screen key.