EasyManua.ls Logo

Waters Alliance e2695 - Page 65

Waters Alliance e2695
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Connect tubing and attachments
July 29, 2013, 715003794 Rev. B 65
Three-column selection valve configuration (as shipped):
Install the 3-column selection valve after installing the column heater or
column heater/cooler (see “Install the column heater and column
heater/cooler” on page 81).
Required material
Open-end wrench, 5/16-inch
To connect the 3-column selection valve to the fluid path of the sample
management system:
1. Use the 5/16-inch wrench to connect the red line (from the e2695
injector) to the IN port on the column selection valve.
2. Use the tubing and fittings supplied with the 3-column selection valve to
connect the OUT port of the column selection valve to the detector inlet.
1
2
3
1
2
3
Column 1
Restrictor
OUT
IN
Waste
Out to
detector
Red line
in

Table of Contents

Related product manuals