2 Safety
Contra-indications
Defibrillation may not be carried out on patients with a body
temperature < 27 °C.
Defibrillation may not be carried out on patients suffering
ventricular tachycardia with a pulse.
Cardioversion may not be carried out on patients who have
overdosed on digitalis.
2.2 Intended areas of application
Mobile use in emergency medicine or primary care in
emergency response
Transport between hospital rooms and departments
Transport between hospitals and other premises by
ambulance, aircraft, helicopter or ship
2.3 Possible side effects/complications
Electrostimulation may cause burns, skin irritation and
arrhythmias.
Electrostimulation may damage implanted medical devices or
those connected to the patient externally.