12 Configuration by the operator
12.6 Pairing the device with an external
device via the Bluetooth
®
interface
(only with Bluetooth
®
data
transmission option)
Requirement The Bluetooth
®
data transmission option is enabled and
activated (see “13.12 System settings”, page 342).
An external device with a Bluetooth
®
interface (Example:
System for digital patient data recording) is within range.
1. Activate the operator menu (see “13.1 Activating the operator
menu”, page 307).
2. Select Communication settings | Pair Bluetooth
®
devices.
The device waits to pair with an external device.
3. Activate the Bluetooth
®
interface of the external device.
4. Check whether the PIN displayed on the device is shown on the
external device.
5. If the PIN displayed on the device and on the external device are
identical: Confirm the pairing on the device and the external
device.
Device and external device are paired.
Result The device and the external device are paired via the Bluetooth
®
interface. The external device can connect to the device and
interrogate device data via the Bluetooth
®
interface.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals