18 Technical data
18.15 Alarm delay times
Alarm Delay time
Battery defective 10.0 s
Insert battery 10.0 s
Asystole
Patient is connected with this condition
Patient already connected
9.6 s
8.5 s
ECG module defective 10.0 s
Heart rate q (with derivation via defibrillation
electrodes)
Patient is connected with this condition
Patient already connected
3.0 s
1.7 s
Heart rate r (with derivation via defibrillation
electrodes)
Patient is connected with this condition
Patient already connected
9.7 s
6.0 s
Heart rate q (with derivation via ECG electrodes)
Patient is connected with this condition
Patient already connected
8.0 s
2.0 s
Heart rate r (with derivation via ECG electrodes)
Patient is connected with this condition
Patient already connected
14.9 s
5.0 s
Pulse rate q 10.0 s
Pulse rate r 10.0 s
SpO
2
saturation q 10.0 s
SpO
2
saturation r 10.0 s
VF/VT
Patient is connected with this condition
Patient already connected
11.3 s
5.3 s