EasyManua.ls Logo

Weinmann MEDUCORE Standard2 - Transferring the Device Configuration to Another Device

Weinmann MEDUCORE Standard2
396 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WM 68401 04/2021
MEDUCORE Standard
2
EN 303
12 Configuration by the operator
12.7 Transferring the device configuration
to another device
Using the device’s SD card, you can transfer device settings saved
in the operator menu to another device.
1. Activate the operator menu (see “13.1 Activating the operator
menu”, page 307).
2. Select System settings | SD card | Export device
configuration to SD card.
3. Confirm with the OK function button.
The export process starts.
4. Remove the SD card (see “4.6.2 Removing the SD card”,
page 93).
5. Insert the SD card into another device (see “4.6.1 Inserting an
SD card”, page 92).
6. Activate the operator menu of the other device (see “13.1
Activating the operator menu”, page 307).
7. Select System settings | SD card | Import device
configuration from SD card.
8. Confirm with the OK function button.
The import process starts.
Result The device settings have been transferred to another device.
Risk of injury from different alarm presets in the same or
similar devices!
Different alarm presets in the same or similar devices in different
application areas may confuse the user and result in injury to the
patient.
Select the same alarm presets in the same or similar devices.

Table of Contents