13 Operator menu
13.4 Alarm settings
Start analysis
automatically
-
Activated
Deactivated
Here you can set whether
cardiac rhythm analysis starts
automatically, or manually at
the touch of a button.
Activated
Pictograms in
parameter view
-
Activated
Deactivated
Here you can set whether
pictograms instead of
parameters are displayed in
parameter view in AED
mode.
Deactivated
Charging during
analysis
-
Activated
Deactivated
Here you can set whether the
shock capacitor is to be
charged at the same time as
cardiac rhythm analysis, and
so independently of the
analysis result. If this setting
is deactivated, the service life
of the shock capacitor is
extended.
Activated
Parameter
Possible
values
Description
Factory
setting
Adult AED settings/Child AED settings
Risk of injury due to alarm limits which are too high or too
low!
Alarm limits which are either too high or too low may prevent the
device triggering an alarm, thereby putting the patient at risk.
Always set alarm limits which have been adapted to the
patient.
Risk of injury from different alarm presets in the same or
similar devices!
Different alarm presets in the same or similar devices in different
application areas may confuse the user and result in injury to the
patient.
Select the same alarm presets in the same or similar devices.