250 EN MEDUCORE Standard
2
WM 68401 04/2021
9 Application menu
9.3 Settings
Parameter
Possible
values
Description Factory setting
Metronome (only
in manual and AED
modes)
Off
15:2 (only in
manual and
AED modes,
child)
30:2 (only in
manual and
AED modes,
adult)
Here you can set the metronome rhythm.
AED mode adult:
30:2
AED mode child:
15:2
Manual mode:
off
SYNC (only with
Cardioversion
option, only in
manual mode)
Activated
Deactivated
Here you can set whether shocks are to be
delivered synchronized. Once a shock has
been delivered, then depending on the
settings in the operator menu (Operator
menu | Manual mode settings |
Defibrillation after cardioversion)
another cardioversion or defibrillation can be
carried out.
Deactivated
12-lead ECG (only
with 12-lead ECG
option)
-
Here you activate the 12-lead ECG function
mode. If you start 12-lead ECG function
mode from AED or manual mode, the device
automatically switches to monitor mode in
the background.
-
Print (only with
Printing option)
-
Here you can print the following printouts:
Live printout
12-lead ECG printout
Replay printout
The Printing menu item is grayed out if the
printer is not connected. When the printer is
printing, the menu item is called Stop print
xx s. A printing process can be finished
prematurely by pressing this menu item.
A counter indicates the time remaining until
the end of printing.
-