236 EN MEDUCORE Standard
2
WM 68401 04/2021
8 Hygienic reprocessing
8.4 Disinfecting parts by wiping
Requirement Hygienic reprocessing is prepared (see “8.2 Preparing hygienic
reprocessing”, page 232).
The parts have been cleaned manually and look clean (see “8.3
Cleaning parts manually”, page 233).
1. For parts approved for disinfection by wiping, refer to the
cleaning and disinfection plan (see “8.7 Cleaning and
disinfection plan”, page 240).
2. For the agents, dose, and exposure time for the individual
parts, refer to the cleaning and disinfection plan.
3. Prepare the solution for disinfection as specified by the
disinfectant manufacturer.
4. Disinfect parts by wiping with one of the products listed.
When doing so, please note:
Uneven surfaces and grooves (e.g. navigation knob) must
be adequately wetted with disinfectant.
The exposure time specified by the manufacturer of the
disinfectant must be observed.
The inner surfaces of MCS2-SoftTip pulse oximetry sensors
must be turned inside out for cleaning.
The attachments for large electrode surfaces (adults) must
be left on the paddles.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals