WM 68401 04/2021
MEDUCORE Standard
2
EN 235
8 Hygienic reprocessing
6. Surround cables/tubes firmly with the cloth and draw them
through the cloth so that they are completely wetted.
When doing so, please note: Stretch out coiled cables.
7. If visible dirt is still present: Repeat manual cleaning.
8. Wipe down parts with a damp cloth to remove residues of the
cleaning agent.
9. Allow parts to dry completely at room temperature.
10. If required: Dry parts manually with a lint-free cloth.
Result Parts have been cleaned manually.