MEDUCORE Standard
2
EN 83
4 Preparation
4.2.2 Stowing components and accessories on the
portable unit
Follow the instructions for use of the portable unit.
4.3 Connecting a power supply
1. Check battery status (see “3.5 Battery and battery status
indicator”, page 57).
2. If necessary: Charge battery (see “4.4.2 Charging the battery
in the device”, page 86).
Risk of injury due to missing battery!
Operation with line power without a battery prevents the device
being fully ready for use as the shock capacitor in the device
cannot charge. This prevents shock delivery and delays the
patient’s treatment.
Only operate the device with the battery inserted.
Risk of injury from battery being incorrectly inserted!
An incorrectly inserted battery may lead to device failure and thus
injure the patient.
Always push the battery into the battery compartment until
you hear it engage and it is flush with the housing.