WM 68401 04/2021
MEDUCORE Standard
2
EN 299
12 Configuration by the operator
7. When registration is successful: Confirm with OK.
Result The device has been registered with WEINMANN Connect. The
device can upload data to WEINMANN Connect with immediate
effect.
12.5 Configuring e-mail delivery 12-lead
ECG (only with E-mail delivery 12-lead
ECG option)
Requirement The 12-lead ECG option is enabled and activated (see “13.12
System settings”, page 342).
The E-mail delivery 12-lead ECG option is enabled and
activated (see “13.12 System settings”, page 342).
A default network has been configured (see “12.3 Configuring
default network”, page 292).
Risk of injury due to delayed delivery!
The user can deliver a 12-lead ECG to a teleconsultation service for
further diagnosis. Delivering a 12-lead ECG is purely to obtain a
second opinion, and does not replace adequately trained
specialists at the session location. Specialist staff at the session
location are responsible for treatment within the scope of their
training and in accordance with the specifications of the operator.
If the 12-lead ECG reaches the teleconsultation service very late or
not at all for technical reasons, delayed treatment may result in a
potentially life-threatening situation. The operator is responsible
for ensuring that the necessary configuration for 12-lead ECG
delivery has been carried out correctly.
When configuring e-mail delivery 12-lead ECG:
Test e-mail delivery.
If the IT infrastructure permits it: Configure several networks
for various devices or session locations so that if one network
is unavailable, it is possible to select another network.
Check the configuration of the networks at regular intervals
and update them if necessary.
Ensure that the e-mail addresses are correct and complete.
Check e-mail addresses at regular intervals and update them if
necessary.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals