18 Technical data
18.4 Master cable
18.5 Battery
18.6 Power supply unit and charger
Specification Master cable
Service life 5000 connector/connection cycles
Specification Battery
Type Li-ion
Dimensions (W x H x D) 97 mm x 127 mm x 33 mm
Weight 450 g
Shock capacity (at 20 °C with new and fully charged
battery)
350 shocks at 200 J
Monitoring capacity Approx. 5 h
Battery operating time for resuscitation 4 h with shocks of 200 J
Rated capacity 4.2 Ah (45.6 Wh/46.4 Wh/46.8 Wh)
Rated voltage 10.8 V
Charging time (0 % to 90 %) Approx. 3.5 h
Charging temperature 0 °C to +45 °C
Service life At least 300 cycles or 2 years
Storage Storage above 60 °C maximum 1 week
Specification Power supply unit and charger
Manufacturer PROTEK POWER
Type PMP120F-13-K24
Operation:
Temperature range
Humidity
•Air pressure
Altitude above mean sea level
0 °C to +40 °C
5 % rh to 95 % rh, no condensation
700 hPa to 1100 hPa
-500 m to 3000 m
Input current 1.4 A to 0.6 A
Input voltage 100 V-240 V~/50 Hz-60 Hz
Rated voltage output 15 V
Disconnection from line power
Pulling out the power plug disconnects the device
from line power on all poles.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals