268 EN MEDUCORE Standard
2
WM 68401 04/2021
11 Alarms and faults
11 Alarms and faults
11.1 General instructions
The device emits an alarm as long as the cause continues to exist.
Once the cause of the alarm no longer exists, the device no longer
emits the alarm. Some alarms are excepted from this. The duration
of these alarms is marked in the tables.
The device emits physiological and technical alarms. Every alarm
has a certain priority:
If more than one alarm is active, the device handles this as follows:
• Multiple alarms of different priorities: The device displays the
alarm with the highest priority. Alarms with a lower priority do
not appear until the higher-priority alarms are no longer active.
• Multiple alarms of identical priorities: The device displays the
alarms in turn.
Priority
Color in
alarm line
Meaning
High priority Red
High-priority alarms warn of imminent
fatal or irreversible patient injuries or
of device faults.
Medium priority Yellow
Medium-priority alarms warn of
immediate reversible patient injuries
or of minor device faults.
Low priority Turquoise
Low-priority alarms warn of delayed
minor injuries or inconvenience to the
patient or minor restrictions on the
device.