374 EN MEDUCORE Standard
2
WM 68401 04/2021
18 Technical data
18.9 CARDIObiphasic defibrillation system
Specification Defibrillation system
Operating mode
Semi-automatic (AED mode)
Manual (manual mode)
Shock form Biphasic, current-limited, impedance-compensated
Patient impedance:
Maximum
Minimum
200 Ω
5 Ω
Shock sequence
Adjustable:
Energy progression:
• First shock: 1 J to 200 J
• Further shocks: 1 J to 200 J
Analysis time 8 s under typical conditions
Duration of analysis and shock charging in
AED mode:
With fully charged source of energy
After 15 discharges
After 6 discharges
8 s to 15 s
8 s to 15 s
8 s to 15 s
Duration of switching on device, analysis, and shock
charging in AED mode:
With fully charged source of energy
After 15 discharges
After 6 discharges
< 25 s
< 25 s
< 25 s
CPR phase adjustable 120 s to 300 s
Energy level adjustable
1 J to 200 J (see “18.18 The CARDIObiphasic shock
pulse”, page 386)
Duration of shock charging in manual mode 2 s to 9 s (depending on the selected shock energy)
Impedance compensation Yes
Defibrillator voltage 2 kV
Recovery time for derivation of defibrillation
electrodes after a defibrillation shock
5 s
Analysis unit
Shock capacitor charging is canceled in AED mode if
a shockable signal was detected during the analysis,
but a non-shockable signal is detected during
subsequent charging.