WM 68401 04/2021
MEDUCORE Standard
2
EN 271
11 Alarms and faults
11.2 Alarm messages
11.2.1 High-priority alarm (red)
Alarm Cause Remedy
Asystole Asystole detected
Act in accordance with currently
applicable guidelines.
If medically indicated, carry out
suitable treatment.
Battery defective Battery defective
Let the device run on battery
power without line power until it
switches off. Fully recharge battery
(see 4.4.2, p. 86). If the device
continues to display the alarm:
Change battery (see 4.4.4, p. 88).
Battery empty Very low battery status
Change battery (see 4.4.4, p. 88).
Connect device to line power (see
4.3, p. 83) and charge battery (see
4.4.2, p. 86).
Check defibrillation
electrodes
(Manual mode)
Defibrillation electrodes no longer
attached/not attached correctly to
the patient
Attach defibrillation electrodes to
the patient correctly (see 6.5.1, p.
122).
Check patient contact
Paddles are not in contact/not
correctly in contact with the
patient
Check contact between paddles
and patient. If the alarm persists:
Cancel shock delivery.
Defibrillation module
defective
Internal defibrillation module
defective
Have the device repaired.
Insert battery
(Manual mode)
Battery not inserted/not inserted
correctly
Insert battery correctly (see 4.3, p.
83).
Plug in Pad connector
(Manual mode)
Defibrillation electrodes not
connected/not connected correctly
to the master cable
Connect defibrillation electrodes
to the master cable correctly (see
6.5.1, p. 122).
Paddles not connected/not
connected correctly to the master
cable
Connect paddles to the master
cable correctly (see 6.5.2, p. 126).
Master cable not connected/not
connected correctly to the device
Connect the master cable to the
device correctly (see 6.5.1, p. 122).
Shock not successful
(Manual mode)
No patient contact during shock
delivery
Wait 10 seconds.
Charge the shock capacitor again
using the Charge function button.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals