164 EN MEDUCORE Standard
2
WM 68401 04/2021
6 Operation
5. If necessary: Make the following SpO
2
settings in the user
menu (see “10.3.4 SpO
2
settings”, page 261):
• Adapt the feed rate of the plethysmogram.
• Give audio pulse tone output priority over heart rate tone
output so that the tone level is dependent on oxygen
saturation even when the ECG cable or defibrillation
electrodes are connected to the patient.
6. If necessary: Set alarm limits in the user menu (see “10.3.1
Alarm settings”, page 255).
Alternatively:
Set automatic alarm limits in the application menu (see “9
Application menu”, page 248).
7. If necessary: Deactivate the pulse tone with the heart rate tone/
pulse tone function button .
8. If there are artifacts in the SpO
2
curve or if signal quality is poor
(bar in the SpO
2
parameter field): Reposition the pulse
oximetry sensor on the patient’s limb.
Result Pulse oximetry monitoring is carried out.
6.10 6-lead ECG monitoring
6.10.1 Preparing 6-lead ECG monitoring
Requirement • The device is switched on (see “6.1 Switching on the device”,
page 118).
• A patient group is selected (see “6.4 Selecting patient group”,
page 120).
Risk of injury due to live connection for 12-lead ECG extension
cable!
The connection for the 12-lead ECG extension cable may be live
and lead to an electric shock if contacted. This may injure the user.
Always seal the 12-lead ECG extension cable connection with
the protective cap if no 12-lead ECG extension cable is
connected.