EasyManuals Logo

Weinmann MEDUCORE Standard2 User Manual

Weinmann MEDUCORE Standard2
396 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #367 background imageLoading...
Page #367 background image
WM 68401 04/2021
MEDUCORE Standard
2
EN 367
18 Technical data
18 Technical data
18.1 Device
Risk of injury due to ignoring the technical data of
components and accessories!
The technical data of components and accessories may differ from
device technical data. Ignoring the technical data of components
and accessories may injure the patient, the user, and bystanders.
Observe the technical data of components and accessories.
Follow the instructions for use of the components and
accessories.
Specification Device
Product class according to Directive
93/42/EEC
IIb
Dimensions (W x H x D) 242 mm x 137 mm x 130 mm
Weight:
Without battery
With battery
2.25 kg
2.75 kg
Continuous operation:
Temperature range
Humidity
•Air pressure
Altitude above mean sea level
Unless stated otherwise, these
conditions also apply to all components
and all accessories.
0 °C to +40 °C
15 % rh to 95 % rh, no condensation
540 hPa to 1100 hPa
-500 m to 5000 m
Transient operation (temporary
operation at the session location):
Temperature range
Humidity
•Air pressure
Altitude above mean sea level
Unless stated otherwise, these
conditions also apply to all components
and all accessories.
-20 °C to +55 °C
15 % rh to 95 % rh, no condensation
540 hPa to 1100 hPa
-500 m to 5000 m

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals