368 EN MEDUCORE Standard
2
WM 68401 04/2021
18 Technical data
Storage/transport (see “15 Storage”,
page 354):
• Temperature range
• Humidity
•Air pressure
• Altitude above mean sea level
Unless stated otherwise, these
conditions also apply to all components
and all accessories.
-40 °C to +70 °C
15 % rh to 95 % rh, no condensation
540 hPa to 1100 hPa
-500 m to 5000 m
Electrical connection (rated voltage) 12 V to 15.1 V
Max. power consumption 30 W
Disconnection from line power
Pulling out the power plug disconnects the device from line power
on all poles.
Operation with on-board power supply:
• Rated voltage
• Max. internal resistance of on-
board power supply
• Maximum current consumption
12 V
500 mΩ
< 3 A
Operating mode Continuous duty
Classification acc. to EN 60601-1:
• Type of protection against elec.
shock
• Degree of protection against elec.
shock (pulse oximetry sensor)
• Degree of protection against elec.
shock (ECG electrodes)
• Degree of protection against elec.
shock (defibrillation electrodes/
paddles)
• Degree of protection against elec.
shock (NIBP cuff)
Protection class II
Degree of protection BF - defibrillation-proof
Degree of protection CF - defibrillation-proof
Degree of protection BF - defibrillation-proof
Degree of protection BF - defibrillation-proof
Degree of protection against
• Ingress of solid objects
• Ingress of dust
• Ingress of water with harmful
effect
IP55: Protected against dust and water jets
Mechanical resistance to falls 1 m
Electromagnetic compatibility (EMC)
acc. to EN 60601-1-2:
Radio interference suppression
Radio interference immunity
Test parameters and limit values can be requested from the
manufacturer WEINMANN Emergency if required.
EN 55011
EN 61000-4 (parts 2 to 6, 8, and 11)
Specification Device