EasyManuals Logo

Weinmann MEDUCORE Standard2 User Manual

Weinmann MEDUCORE Standard2
396 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #380 background imageLoading...
Page #380 background image
380 EN MEDUCORE Standard
2
WM 68401 04/2021
18 Technical data
18.13 Non-invasive blood pressure
measurement (NIBP monitoring)
18.14 Operation/data management
Specification
Non-invasive blood pressure
measurement
Manufacturer SunTech Medical
Measuring method
Oscillometric, diastolic values correspond to phase 5
Korotkoff sounds
Measurement range (systolic):
Adults
•Children
•Infants
40 mmHg to 260 mmHg
40 mmHg to 230 mmHg
40 mmHg to 130 mmHg
Measurement range (diastolic):
Adults
•Children
•Infants
20 mmHg to 200 mmHg
20 mmHg to 160 mmHg
20 mmHg to 100 mmHg
Accuracy (during operation between 0 °C and 50 °C) ±3 mmHg
Specification Operation/data management
Display
Illuminated symbols
Device status indicators
Audio output
•Voice prompts
•Alarm tones
Audible signals
Session documentation
Automatic recording of measured values for ECG,
SpO
2
and NIBP and of event data
Data transmission via SD card SD card with 32 GB
Data transmission via WiFi
Data transfer types: 12-lead ECG, function check
results, session data, service data
WiFi: 802.11a/b/g/n
Security: WPA2-PSK
Protocol: https
Port: 443
Data transmission via Bluetooth
®
interface Data transfer type: Session data
Data evaluation Via DEFIview PC software

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish