EasyManua.ls Logo

Weinmann MEDUCORE Standard2 - Carrying Out a Function Check

Weinmann MEDUCORE Standard2
396 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
104 EN MEDUCORE Standard
2
WM 68401 04/2021
5 Function check
5.1 Intervals
5.2 Carrying out a function check
Requirement • The device is disconnected from the patient.
• A battery with a battery status of at least 1 LED is inserted in
the device.
Alternatively:
The display shows at least battery status .
• The SD card is in the SD card slot.
• The master cable is connected.
• The pulse oximetry sensor connecting cable with pulse
oximetry sensor is connected.
1. Check the following parts for external damage:
• Device
• Defibrillation electrode packaging
• Master cable
• Paddles
•ECG cable
• Pulse oximetry sensor connecting cable
• Pulse oximetry sensor
• NIBP connecting tube
•NIBP cuff
•Printer
Part concerned Interval
Device including accessories
• Before every use
• After every hygienic reprocessing
• After every repair
• After every software update

Table of Contents