3 Description
3.10 Options
You can tailor the range of functions on the device to your needs
using the options. You can enable these options using a device-
specific enable code (see “12.1 Enabling options”, page 290). The
following options are available:
Option Description Parts required/conditions
Manual defibrillation This option enables manual mode. -
Cardioversion
This option allows cardioversion to
be carried out in manual mode.
Manual defibrillation option
enabled and activated
Printing
This option allows various reports
and printouts to be printed:
Function check report
Live printout
Defibrillation report
12-lead ECG printout (only with
12-lead ECG option)
Replay printout (only with
Replay view option)
Session report
•Printer
Printer paper
Printer battery
Charging station for printer
battery with power supply
unit and charger
•Printer case
12-lead ECG This option enables 12-lead ECG.
ECG cable with connection
for 12-lead ECG extension
cable
12-lead ECG extension cable
E-mail delivery 12-lead ECG
This option allows you to deliver a
12-lead ECG to an e-mail address.
12-lead ECG option enabled
and activated
Network connected to the
Internet
Device registered with
WEINMANN Connect (see
“12.4 Registering the device
with WEINMANN Connect”,
page 298)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals