278 EN MEDUCORE Standard
2
WM 68401 04/2021
11 Alarms and faults
* If the device outputs this alarm, it will no longer output a
simultaneously occurring SpO
2
measured value alarm. If signal
quality drops to < 20 %, the device will also stop displaying the
SpO
2
measured value and the plethysmogram.
11.3 Faults
If you are unable to remedy faults with the aid of the table, you
should contact the manufacturer WEINMANN Emergency, or
technicians expressly authorized by WEINMANN Emergency, to
have the device repaired. To avoid more serious damage, do not
continue using the device.
SD card defective SD card write/read error
Insert new SD card (see 4.6.1, p.
92), switch off device (see 6.2, p.
119), and switch device back on
after at least 30 s (see 6.1, p. 118).
Resume session and do not record
any more data.
SD card full SD card full
Insert new SD card (see 4.6.1, p.
92), switch off device (see 6.2, p.
119), and switch device back on
after at least 30 s (see 6.1, p. 118).
Erase data on current SD card.
SpO
2
sensor defective
Pulse oximetry sensor connecting
cable defective
Replace pulse oximetry sensor
connecting cable.
Pulse oximetry sensor defective Replace pulse oximetry sensor.
SpO
2
signal quality r *SpO
2
signal quality < 40 %
Attach pulse oximetry sensor to
the patient correctly (see 6.9.1, p.
157).
If medically indicated, carry out
suitable treatment.
Update operator menu access
code
(Alarm deactivated after 10 s)
Access code for the operator menu
is 000000 (status when device
supplied)
Wait 10 seconds.
Change operator menu access
code (see “13.12 System
settings”, page 342).
Alarm Cause Remedy

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals