11 Alarms and faults
11.2.3 Low-priority alarm (turquoise)
Alarm Cause Remedy
Battery operation
(Alarm deactivated after 10 s)
Line power too low
Restore line power.
Line power disconnected by
removing from the wall mounting
Power outage
Check seating of SpO
2
sensor
Pulse oximetry sensor not
attached/not attached correctly to
the patient
Attach pulse oximetry sensor to
the patient correctly (see 6.9.1, p.
157).
Erroneous date
Date read-out error
Insert battery (see 4.3, p. 83).
Reset date (see 10.3.7, p. 264).
Internal module defective
If this recurs: Have the device
repaired.
Real Time Clock battery defective
or empty
Have the device repaired.
Insert battery
(Monitor mode)
Battery not inserted/not inserted
correctly
Insert battery correctly (see 4.3, p.
83).
Insert SD card No SD card present/detected
Insert new SD card (see 4.6.1, p.
92), switch off device (see 6.2, p.
119), and switch device back on
after at least 30 s (see 6.1, p. 118).
Internal memory full
No SD card inserted and session
duration > 4 h-10 h
To continue to record the session:
Insert SD card (see 4.6.1, p. 92),
copy data to SD card, and resume
session with SD card.
Resume session and do not record
any more data.
Plug in SpO
2
sensor
SpO
2
connector of the pulse
oximetry sensor connecting cable
not connected/not connected
correctly to the device
Connect SpO
2
connector of the
pulse oximetry sensor connecting
cable to the device correctly.
Printer battery empty Status of printer battery very low
Change printer battery (see 4.7.2,
p. 97).
Charge printer battery (see 4.7.1,
p. 94).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals