WM 68401 04/2021
MEDUCORE Standard
2
EN 97
4 Preparation
6. When the LED on the power supply unit and charger is green:
Take the printer battery out of the charging station for the
printer battery.
Alternatively:
If the FULL CHARGED LED is green: Take the printer battery out
of the quadruple charging station for the printer battery.
Result The printer battery is fully charged.
4.7.2 Inserting the printer battery
Requirement The printer battery is fully charged (see “4.7.1 Charging the
printer battery”, page 94).
The printer is switched off.
1. Turn over the printer.
2. Insert the printer battery in the battery compartment of the
printer until you hear it engage.
Risk of injury from electric shock when simultaneously
touching a printer and the patient!
Simultaneously touching a printer and the patient may lead to an
electric shock and injure the patient.
Stand clear of the patient if the printer is out of its case.