8 Hygienic reprocessing
8.5 Subjecting bags to hygienic
reprocessing
Requirement Hygienic reprocessing is prepared (see “8.2 Preparing hygienic
reprocessing”, page 232).
1. For the agents, dose, and exposure time for the bags, refer to
the cleaning and disinfection plan.
2. Soak the bag in mains water for at least 10 minutes.
3. Brush contaminated areas under running mains water for at
least 2 minutes using a cleaning brush until such areas look
clean.
4. Rinse off cleaned areas under running mains water for
2 minutes.
5. Wash the open bag in a washing machine with cleaning agent
and disinfectant but without a spin cycle:
Prewash: 5 min at 20 °C
Main wash: 30 min at 30 °C
Result A bag has been hygienically reprocessed.
8.6 Preparing parts for reuse
Requirement The parts have been subjected to hygienic reprocessing according
to the cleaning and disinfection plan.
1. Check all parts for damage resulting from use (e.g. tension
cracks or cable breaks).
2. Replace damaged parts.
3. When the device has been disassembled from the protective
transport bag and the retaining plate: Mount the device in the
protective transport bag (see “4.1.1 Mounting the device in the
protective transport bag”, page 74).
Alternatively:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals