8 Hygienic reprocessing
3. Disassemble the pulse oximetry sensor and pulse oximetry
sensor connecting cable (see “7.3 Disassembling pulse
oximetry sensor and pulse oximetry sensor connecting cable”,
page 222).
4. Disassemble the ECG cable (see “7.4 Disassembling ECG
cable”, page 224).
5. Disassemble the NIBP cuff and NIBP connecting tube (see “7.5
Disassembling the NIBP cuff and NIBP connecting tube”,
page 225).
6. Remove the remaining accessories from the device.
7. In the event of contamination or suspected contamination of
inaccessible areas of the device, protective transport bag or
portable unit: Disassemble the device from the protective
transport bag and retaining plate (see “7.6 Disassembling the
device from the protective transport bag and the retaining
plate”, page 227).
Alternatively:
Disassemble the device from the portable unit (see “7.7
Disassembling the device from the portable unit”, page 229).
When doing so, please note: Disassembly is described in the
instructions for use of the portable unit.
8. Dispose of all disposables properly (see “16 Disposal”,
page 357).
Result All parts have been prepared for hygienic reprocessing.
8.3 Cleaning parts manually
Requirement The parts exhibit visible dirt.
Hygienic reprocessing is prepared (see “8.2 Preparing hygienic
reprocessing”, page 232).
1. For parts approved for cleaning, refer to the cleaning and
disinfection plan (see “8.7 Cleaning and disinfection plan”,
page 240).
2. For the agents, dose, and exposure time for the individual
parts, refer to the cleaning and disinfection plan.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals