234 EN MEDUCORE Standard
2
WM 68401 04/2021
8 Hygienic reprocessing
3. Prepare the cleaning solution as specified by the cleaning agent
manufacturer.
4. To remove all visible dirt: Brush parts thoroughly using a
commercially available soft brush suitable for plastic and
wetted with the cleaning agent.
When doing so, please note: Uneven surfaces and grooves
(e.g. navigation knob) must be brushed especially thoroughly.
5. Wipe over parts with a clean, lint-free cloth moistened with
cleaning solution until they look clean.
When doing so, please note:
Use a new cloth for every cleaning process.
All surfaces must be wiped carefully.
All surfaces must be wetted with cleaning solution.
The exposure time specified in the cleaning and disinfection
plan must be observed.
Uneven surfaces and grooves need wiping over especially
carefully.
The inner surfaces of MCS2-SoftTip pulse oximetry sensors
must be turned inside out for cleaning.
The attachments for large electrode surfaces (adults) must
be left on the paddles.
The attachments for large electrode surfaces must be
removed to clean the attachments for small electrode
surfaces (children/infants).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals