270 EN MEDUCORE Standard
2
WM 68401 04/2021
11 Alarms and faults
The device additionally displays physiological alarms by means of
the respective parameter field flashing.
The device’s alarm behavior depends on the mode selected:
In AED mode, the device does not emit alarms. The device
provides information by means of AED instructions and voice
prompts about technical conditions which lead to AED mode
being impaired.
Audio alarm output is deactivated in manual mode. It can be
reactivated at any time by pressing the alarm button. In the
operator menu, the operator can specify whether audio alarm
output is to be activated when manual mode is entered.
Audio alarm
Number of pulses 5 3 1
Duration of
pulses
120 ms 200 ms 200 ms
Pulse interval 240 ms 200 ms Not applicable
Pulse frequency 480 Hz 840 Hz 1200 Hz
Alarm light
Color Red Off Off
Frequency 2 Hz Not applicable Not applicable
Signal ratio 33 % on Not applicable Not applicable
Size 10 mm x 35 mm
Type of alarm
signal
High priority
Medium
priority
Low priority

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals