WM 68401 04/2021
MEDUCORE Standard
2
EN 313
13 Operator menu
Audio outputs
CPR voice
prompts
Activated
Deactivated
Here you can set whether the
voice prompts for guiding
chest compressions and for
ventilation are to be given.
The voice prompt Ventilate
twice is only given if the
metronome is set to the
rhythm 15:2/30:2.
Deactivated
Metronome
off
15:2/30:2
Here you can set the
metronome rhythm.
15:2/30:2
CV ratio
30:2
15:2
Here you can set the
compression/ventilation
ratio.
•15:2: After
15 metronome beats,
there is a pause to allow
for ventilation.
•30:2: After
30 metronome beats
there is a pause to allow
for ventilation.
•AED mode
adult: 30:2
•AED mode
child: 15:2
Metronome
frequency
100/min to
120/min
Here you can set the
metronome frequency.
100/min
Parameter
Possible
values
Description
Factory
setting
Adult AED settings/Child AED settings