312 EN MEDUCORE Standard
2
WM 68401 04/2021
13 Operator menu
13.3.2 Possible values
Parameter
Possible
values
Description
Factory
setting
Adult AED settings/Child AED settings
Energy
progression
First shock
1 J to 200 J
Here you can set the shock
energy of the first shock in
AED mode.
•AED mode
adult: 150 J
•AED mode
child: 75 J
Further shocks
Here you can set the shock
energy of further shocks in
AED mode.
•AED mode
adult: 200 J
•AED mode
child: 75 J
CPR phase
Duration 120 s to 300 s
Here you can set the time
interval between cardiac
rhythm analyses.
The device only outputs
certain AED instructions and
voice prompts after CPR is
complete, as complete and
correct carrying out of CPR
takes priority.
120 s
Pause for
ventilation
3 s to 8 s
Here you can set the time
interval for ventilation
between chest compressions.
When choosing a setting,
consider whether voice
prompts are deactivated:
• When voice prompts are
activated, the pause
begins after the last
metronome sound and
ends when the voice
prompts begin.
• When voice prompts are
deactivated, the pause
begins after the last
metronome sound and
ends when the first
metronome sound
begins.
5 s