EasyManuals Logo

Weinmann MEDUCORE Standard2 User Manual

Weinmann MEDUCORE Standard2
396 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #123 background imageLoading...
Page #123 background image
WM 68401 04/2021
MEDUCORE Standard
2
EN 123
6 Operation
1. Connect the master cable connector to the Pad connection for
master cable on the device.
2. Select defibrillation electrodes suitable for adults (Adult) or
children (Pediatric).
3. In the case of defibrillation electrodes for children (Pediatric):
Tear open the defibrillation electrode packaging and take out
the defibrillation electrodes.
Risk of injury due to incorrectly selected size of defibrillation
electrodes!
If the wrong size of defibrillation electrodes is selected, this may
result in sub-optimal defibrillation results or in burns.
Select the correct size of defibrillation electrodes in line with
currently applicable guidelines and not based on the weight
specifications given on the packaging.