WM 68401 04/2021
MEDUCORE Standard
2
EN 147
6 Operation
6. If cardioversion is required: Select shock energy using the
Energy function button.
7. Open the application menu using the navigation knob.
8. Select and activate the SYNC menu item.
Manual mode switches to cardioversion:
•The SYNC marking indicates that the next shock delivery
will be synchronized with the R wave.
In the ECG, R waves are marked with a triangle .
9. Wait until the ECG curve has stabilized.
10. Select the feed rate of the ECG so that reliable assessment of R
wave detection is possible.
Delayed therapy due to incorrectly selected feed rate for the
ECG!
An incorrectly selected feed rate may make it harder to assess R
waves, delaying therapy.
Select the feed rate of the ECG so that reliable assessment of R
wave detection is possible.