152 EN MEDUCORE Standard
2
WM 68401 04/2021
6 Operation
6. If necessary: Select another ECG lead using the Lead function
button.
7. If cardioversion is required: Select shock energy using the
Energy function button.
8. Open the application menu using the navigation knob.
9. Select and activate the SYNC menu item.
Manual mode switches to cardioversion:
•The SYNC marking indicates that the next shock delivery
will be synchronized with the R wave.
In the ECG, R waves are marked with a triangle .
Risk of injury due to incorrectly selected shock energy in
children!
If the attachments for small electrode surfaces are used in
combination with too high a shock energy on the paddles, this
may lead to injuries in children/infants.
When using the attachments for small electrode surfaces, set a
maximum shock energy of 100 J for children/infants.