6 Operation
6.16.4 Printing a defibrillation report
Requirement A printer has been connected to the device (see “6.16.1
Connecting the printer to the device”, page 196).
•The Defibrillation report menu item is activated in the
operator menu (see “13.9 Printer settings (only with Printing
option)”, page 332).
1. Carry out defibrillation or cardioversion.
On every shock delivery, the printer prints a defibrillation report
(3 s before and 5 s after the shock delivery, shown in abstract
form here):
Result A defibrillation report has been printed.
6.16.5 Printing a function check report
Requirement A printer has been connected to the device (see “6.16.1
Connecting the printer to the device”, page 196).
1. Carry out a function check (see “5 Function check”, page 103).