2 Safety
After using the defibrillator, check the function of electrical
devices in its vicinity.
Maintain separation distances between the defibrillator and
portable and mobile high-frequency communications devices.
Caution Delay in treatment due to simultaneous voice prompts from
defibrillator and ventilator!
If the defibrillator in AED mode is used in conjunction with a
ventilator (MEDUMAT Easy CPR) which also guides the user
through CPR by means of voice prompts, the simultaneous voice
prompts from the defibrillator and ventilator may confuse the user
and delay treatment.
When using the defibrillator in AED mode and a ventilator at
the same time: Switch off the ventilator voice prompts.
Risk of injury due to incorrectly selected size of defibrillation
electrodes!
If the wrong size of defibrillation electrodes is selected, this may
result in sub-optimal defibrillation results or in burns.
Select the correct size of defibrillation electrodes in line with
currently applicable guidelines and not based on the weight
specifications given on the packaging.
2.6.5 Pulse oximetry monitoring
Warning Risk of injury due to incorrect use of the pulse oximetry
sensor!
Incorrect use of the pulse oximetry sensor may falsify
measurement results and lead to patient injury.
Keep the pulse oximetry sensor away from strong
electromagnetic sources (e.g. electrosurgical devices).
Do not use the pulse oximetry sensor in areas subject to X-ray
(e.g. with MRI devices).
Keep the pulse oximetry sensor away from strong and
fluctuating ambient light (including infrared and UV light). If
necessary: Cover the pulse oximetry sensor.
Avoid strong movement of the pulse oximetry sensor. If
necessary: To relieve strain, loop the pulse oximetry sensor
cable and the pulse oximetry sensor connecting cable and fix to
the patient with a plaster.
Do not attach the pulse oximetry sensor to a limb on which
there is already an NIBP cuff or catheter port.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals