318 EN MEDUCORE Standard
2
WM 68401 04/2021
13 Operator menu
Child alarm limit
settings
Heart rate q
35/min - 250/min,
in increments of 5
Here you can set the upper
( ) and lower ( ) limit
values for the patient group
as of which the device is to
emit an alarm.
According to the values set,
the setting ranges of the
upper and lower limit values
depend on each other as
follows:
• The setting range for the
upper alarm limit ends at
least one setting value
above the set value for
the lower alarm limit.
• The setting range for the
lower alarm limit ends at
least one setting value
below the set value for
the upper alarm limit.
150/min
Heart rate r
30/min - 245/min,
in increments of 5
50/min
Pulse rate q
35/min - 250/min,
in increments of 5
150/min
Pulse rate r
30/min - 245/min,
in increments of 5
50/min
SpO
2
saturation q 66 % - 100 % 100 %
SpO
2
saturation r 65 % - 99 % 85 %
NIBP systolic q
45 mmHg -
230 mmHg,
in increments of 5
145 mmHg
NIBP systolic r
40 mmHg -
225 mmHg,
in increments of 5
75 mmHg
NIBP diastolic q
25 mmHg -
160 mmHg,
in increments of 5
100 mmHg
NIBP diastolic r
20 mmHg -
155 mmHg,
in increments of 5
35 mmHg
Parameter
Possible
values
Description
Factory
setting