EasyManua.ls Logo

Weinmann MEDUCORE Standard2 - Page 323

Weinmann MEDUCORE Standard2
396 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WM 68401 04/2021
MEDUCORE Standard
2
EN 323
13 Operator menu
13.5.2 Possible values
Parameter
Possible
values
Description
Factory
setting
Adult
Shock energy 1 J to 200 J
Here you can set what
defibrillation energy is to be
preset in manual mode.
150 J
Metronome
off
15:2
30:2
Here you can set the
metronome rhythm.
off
Metronome
frequency
100/min to
120/min
Here you can set the
metronome frequency.
100/min
Pause for
ventilation
3 s to 8 s
Here you can set the time
interval for ventilation
between chest compressions.
5 s
Child
Infant
Shock energy 1 J to 200 J
Here you can set what
defibrillation energy is to be
preset in manual mode.
75 J
Metronome
off
15:2
30:2
Here you can set the
metronome rhythm.
off
Metronome
frequency
100/min to
120/min
Here you can set the
metronome frequency.
100/min
Pause for
ventilation
3 s to 8 s
Here you can set the time
interval for ventilation
between chest compressions.
5 s
Defibrillation
after
cardioversion
(only with
Cardioversion
option)
-
Activated
Deactivated
Here you can set whether
defibrillation is to follow
cardioversion (box checked)
or cardioversion is to follow
cardioversion (box
unchecked).
Deactivated

Table of Contents

Related product manuals