WM 68401 04/2021
MEDUCORE Standard
2
EN 323
13 Operator menu
13.5.2 Possible values
Parameter
Possible
values
Description
Factory
setting
Adult
Shock energy 1 J to 200 J
Here you can set what
defibrillation energy is to be
preset in manual mode.
150 J
Metronome
off
15:2
30:2
Here you can set the
metronome rhythm.
off
Metronome
frequency
100/min to
120/min
Here you can set the
metronome frequency.
100/min
Pause for
ventilation
3 s to 8 s
Here you can set the time
interval for ventilation
between chest compressions.
5 s
Child
Infant
Shock energy 1 J to 200 J
Here you can set what
defibrillation energy is to be
preset in manual mode.
75 J
Metronome
off
15:2
30:2
Here you can set the
metronome rhythm.
off
Metronome
frequency
100/min to
120/min
Here you can set the
metronome frequency.
100/min
Pause for
ventilation
3 s to 8 s
Here you can set the time
interval for ventilation
between chest compressions.
5 s
Defibrillation
after
cardioversion
(only with
Cardioversion
option)
-
Activated
Deactivated
Here you can set whether
defibrillation is to follow
cardioversion (box checked)
or cardioversion is to follow
cardioversion (box
unchecked).
Deactivated

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals